| Price | Please send inquiry |
| MOQ | 1 set |
| Delivery Time | 35 days |
| Brand | Leader-t |
| Place of Origin | Chengdu, China |
| Certification | CE, ISO |
| Model Number | LDM-D-300-3000L |
| Packaging Details | customized wooden case/shelf, strong paper box, shrink film, EPE foam |
| Payment Terms | T/T,west union,L/C |
| Supply Ability | 1000 sets/month |
| Place of Origin | Chengdu, China | Packaging Details | customized wooden case/shelf, strong paper box, shrink film, EPE foam |
| Optional functions | Thermal disinfection device, U-shaped small circulation, without dead space, Integrated mute cabinet, U-type small circulation, unit without dead space, Online monitoring of residual chlorine | Pump Brand | Grundfos/ CNP |
| Support and Services | Installation guideline support, Maintenance and repair services, Troubleshooting and diagnostics, Software and firmware updates, Customer support and training, Warranty and replacement services | Technical Parameters | Pressure Vessels, Control Panel, Occupied Area, Power Supply, Total Weight, Membrane Housing, Warranty, Membrane Brand, Flow Rate, Pump Brand |
| Model Number | LDM-D-300-3000L | Supply Ability | 1000 sets/month |
| Certification | CE, ISO | Power Supply | 380V/50Hz/ Customized |
| Brand Name | Leader-t | Payment Terms | T/T,west union,L/C |
| Flow Rate | 2500-3000 LPH | Price | Please send inquiry |
| Delivery Time | 35 days | Executive standard | Manufacturing standard |
| Minimum Order Quantity | 1 set | Quality of produced water | Conforms to the requirements of YY0572-2015 Water for Haemodialysis and Related Therapies |
| Item | Specification Description |
|---|---|
| Product Name | Water making equipment for hemodialysis |
| System Water Production Capacity | 300—6000L/H |
| Microorganism Content | Bacteria ≤100 CFU/mL, Endotoxin ≤0.25 EU/mL |
| System Recovery Rate | 50-75% (adjustable) |
| Product Process | Composed of a multi-stage pre-treatment system, a two-stage reverse osmosis system, a disinfection system, a control system, and connecting pipelines. |
| Disinfection Method | Chemical disinfection, pipeline thermal disinfection (optional) |
| Equipment Power | Main unit: 7-13 kVA; Thermal disinfection device: 11-18 kVA |
| Operation Method | PLC + Intelligent display screen |
| Operation Mode | Automatic operation, Manual operation |
| Operating Weight | Approximately 2000-4000 Kg (including filters); Thermal disinfection device: approximately 100 Kg |
| Working Power Supply | AC 380V ±10% (three-phase five-wire system), Frequency: 50Hz ±1Hz |
| Site Requirements | Room area ≥ 20 m², with floor waterproofing treatment, a drainage pipe network, and no direct sunlight exposure. |
| Working Environment | Room temperature 5~40℃, Relative humidity ≤80%RH |
| Raw water requirements | Water supply volume must be at least twice the nominal treatment capacity, with a pressure of 0.15–0.5 MPa and a water temperature of 5–45 °C. |
| Working Pressure | 1.0–1.5 MPa |
| Structural Material | 304/316L stainless steel pipeline, FRP (Fiberglass Reinforced Plastic) storage tank, Medical-grade connector |