n
General
The
model
FLXA®202
2-Wire
Analyzer,
one
model
of
FLEXA®
series,
offers
single
or
dual
sensor
measurement.
The
modular-designed
analyzer
offers
4
kinds
of
measurements
–
pH/ORP
(oxidation-reduction
potential),
contacting
conductivity
(SC),
inductive
conductivity
(ISC)
or
dissolved
oxygen
(DO)
–
with
the
respective
sensor
module.
For
dual
sensor
measurement,
the
combination
of
two
same
type
sensor
inputs
–
pH/ORP
and
pH/ORP
(analog
sensor
only),
SC
and
SC,
and
DO
and
DO
–
are
available
with
two
sensor
modules.
Dual
sensor
measurement
offers
additional
functionalities;
calculated
data
function
and
redundant
system.
Variety
of
calculated
data
from
two
measuring
parameters
is
selectable
for
each
measurement.
On
the
redundant
system
built
on
two
measuring
parameters
of
two
sensor
inputs,
main
output
parameter
is
automatically
switched
over
to
the
second
sensor
output
in
case
of
the
main
sensor’s
failure
condition.
Addition
to
conventional
analog
pH/ORP
sensors,
the
analyzer
FLXA202
can
be
connected
to
Yokogawa’s
digital
sensor,
FU20F
/
FU24F
/
SC25F
Digital
pH/
ORP
SENCOM®
Sensor.
In
the
FLXA202
Human
Machine
Interface
(HMI),
2-wire
type
analyzer
FLXA202
offers
easy
touch
screen
operation
and
simple
menu
structure
in
12
languages.
Menus
of
display,
execution
and
setting
are
displayed
in
a
selected
language.
The
analyzer
FLXA202
automatically
recognizes
the
installed
sensor
module
and
prepares
the
necessary
menus
for
right
configuration,
even
for
dual
sensor
measurement.
For
immediate
measurement,
the
FLXA202
offers
quick
setup
functionality.
The
quick
setup
screen
appears
when
the
analyzer
is
powered.
Only
a
few
setups
–
date/time,
language,
basic
sensor
configurations
and
output
–
will
start
the
measurement.
The
FLXA202
offers
the
best
accuracy
in
measurement
with
temperature
compensation
functionality
and
calibration
functionality.
Sensor
diagnostics
and
sensor
wellness
indication
make
measurement
reliable.
Logbook
of
events
and
diagnostic
data
is
a
useful
information
source
for
maintenance.
For
the
wide
range
of
industrial
environment,
the
FLXA202
is
designed
with
the
enclosure
of
aluminum
alloy
cast
with
corrosion-resistant
coating.
n
Features
•
4
kinds
of
measurements;
pH/ORP,
SC,
ISC
and
DO
•
Dual
sensor
measurement
on
2-wire
type
analyzer;
pH/ORP
and
pH/ORP,
SC
and
SC,
and
DO
and
DO
•
Calculated
data
from
dual
sensor
measurement
•
Redundant
system
on
dual
sensor
measurement
•
Connection
of
FU20F
/
FU24F
/
SC25F
Digital
pH/
ORP
SENCOM
Sensor
•
Easy
touch
screen
operation
on
2-wire
type
analyzer
•
Simple
HMI
menu
structure
in
12
languages
•
Quick
setup
menu
for
immediate
measurement
•
Indication
of
sensor
wellness
•
Enclosure
–
aluminum
alloy
cast.
GS
12A01A03-01EN
©Copyright
Sep.
2015
2nd
Edition
Feb.
02,
2016
2
All
Rights
Reserved.
Copyright
©
2015,
Yokogawa
Electric
Corporation
GS
12A01A03-01EN
n
General
Specifications
1.
Basic
■
Measurement
Object/Sensor
Type
•
pH/Oxidation-reduction
Potential
(pH/ORP)
(analog
sensor)
•
Conductivity
(SC)
•
Inductive
Conductivity
(ISC)
•
Dissolved
Oxygen
(DO)
•
pH/Oxidation-reduction
Potential
(pH/ORP)
(digital
sensor)
Note:
The
available
measurement
object
depends
on
a
sensor
module
installed
on
the
analyzer.
■
Analyzer
Structure
Module
structure
●
Composition
of
Analyzer
One
(1)
Housing
assembly
One
(1)
or
two
(2)
Sensor
modules
●
Combination
of
Sensor
Module
when
two
modules
are
installed
Combinations
of
two
same
sensor
modules
are
available;
pH/ORP
and
pH/ORP
(analog
sensor)
SC
and
SC
DO
and
DO
2.
Measurement
2-1.
pH/Oxidation-reduction
Potential
(pH/ORP)
with
analog
sensors
■
Input
Specification
Dual
high
impedance
input
(≥1012
Ω)
■
Input
Range
pH:
-2
to
16
pH
(with
option
/K:
0
to
14
pH)
ORP:
-1500
to
1500
mV
rH:
0
to
100
rH
Temperature:
Pt1000:
-30
to
140
ºC
Pt100:
-30
to
140
ºC
6.8k:
-30
to
140
ºC
PTC10k:
-30
to
140
ºC
NTC
8k55:
-10
to
120
ºC
3k
Balco:
-30
to
140
ºC
PTC500:
-30
to
140
ºC
■
Output
Range
pH:
min.
span
1
pH
max.
span
20
pH
ORP:
min.
span
100
mV
max.
span
3000
mV
rH:
min.
span
2
rH
max.
span
100
rH
Temperature:
min.
span
25
ºC
max.
span
170
ºC
■
Performance
(Accuracy)
(The
specifications
are
expressed
with
simulated
inputs.)
pH
Linearity:
±0.01
pH
Repeatability:
±0.01
pH
Accuracy:
±0.01
pH
ORP
Linearity:
±1
mV
Repeatability:
±1
mV
Accuracy:
±1
mV
Temperature
with
Pt1000,
6.8k,
PTC10k,
NTC
8k55,
3k
Balco,
PTC500
Linearity:
±0.3
ºC
Repeatability:
±0.1
ºC
Accuracy:
±0.3
ºC
with
Pt100
Linearity:
±0.4
ºC
Repeatability:
±0.1
ºC
Accuracy:
±0.4
ºC
2-2.
Conductivity
(SC)
■
Input
Specification
Two
or
four
electrodes
measurement
with
square
wave
excitation,
using
max
60m
(200ft)
cable
(WU40/
WF10)
and
cell
constants
from
0.005
to
50.0
cm-1
■
Input
Range
Conductivity:
min.:
0
µS/cm
max.:
200
mS
x
(Cell
constant)
(over
range
2000
mS/cm)
Resistivity:
min.:
0.005
kΩ
/
(Cell
constant)
max.:
1000
MΩ
x
cm
Temperature:
Pt1000:
-20
to
250
ºC
Pt100:
-20
to
200
ºC
Ni100:
-20
to
200
ºC
NTC
8k55:
-10
to
120
ºC
Pb36(JIS
NTC
6k):
-20
to
120
ºC
■
Output
Range
Conductivity:
min.
0.01
µS/cm
max.
2000
mS/cm
(max
90%
zero
suppression)
Resistivity:
min.
0.001
kΩ
x
cm
max.
1000
MΩ
x
cm
(max
90%
zero
suppression)
Temperature:
min.
span
25
ºC
max.
span
270
ºC
■
Performance
(Accuracy)
(The
specifications
are
expressed
with
simulated
inputs.)
Conductivity
2
µS
x
K
cm-1
to
200
mS
x
K
cm-1
Accuracy:
±0.5%F.S.
1
µS
x
K
cm-1
to
2
µS
x
K
cm-1
Accuracy:
±1%F.S.
Resistivity
0.005kΩ
/
K
cm-1
to
0.5MΩ
/K
cm-1
Accuracy:
±0.5%F.S.
0.5MΩ
/
K
cm-1
to
1MΩ
/K
cm-1
Accuracy:
±1%F.S.
Temperature
with
Pt1000,
Pb36,
Ni100
Accuracy:
±0.3
ºC
with
Pt100,
NTC
8k55
Accuracy:
±0.4
ºC
Temperature
compensation
NaCl
table:
±1
%
Matrix:
±3
%
Step
response:
90
%
(<
2
decades)
in
7
seconds
Note:
"F.S."
means
maximum
setting
value
of
analyzer
output.
"K"
means
cell
constant.