Price | $25~$60000/Unit |
MOQ | 1unit |
Delivery Time | 10-15days |
Brand | WINLEY |
Place of Origin | CHINA |
Certification | ISO,IEC |
Model Number | HDB-3.15KVA |
Packaging Details | Plywood case |
Payment Terms | T/T |
Supply Ability | 1000units/month |
Usage | Hospital Power | Place of Origin | CHINA |
Phase | Single | Cooling method | Natural or air cooling |
Packaging Details | Plywood case | Model Number | HDB-3.15KVA |
Supply Ability | 1000units/month | Certification | ISO,IEC |
Brand Name | WINLEY | Payment Terms | T/T |
Winding | Enameled Copper Wire | Price | $25~$60000/Unit |
Delivery Time | 10-15days | Efficiency | ≥95% |
Minimum Order Quantity | 1unit | Primary | 220V/120V |
Frequency | 50/60Hz | Coil structure | Toroidal |
Single Phase Medical Isolation Transformer 50/60Hz 220V/120V
1.A
temperature
sensor
is
installed
in
the
winding,
which
can
be
used
to
monitor
the
temperature
of
the
winding.
2.The
whole
transformer
is
treated
with
vacuum
infiltration
paint.
3.Increased mechanical strength and features such as corrosion resistance, low temperature rise, low noise, etc., specially tailored for medical places.
Main technical data:
Parameters: | |
Product Name | Medical isolation transformer |
Product Model | HDB / IT |
Winding Material | Enameled copper wire |
Capacity | 3.15KVA~10KVA |
Input Voltage | 220V/230V/380V/Customized |
Output Voltage | 220V/230V/380V/Customized |
Frequency | 50-60Hz |
Insulation Class | F / H |
Protection Class | IP 00 (construction type: open) / IP23 (enclosure included) |
Withstand Voltage Test |
Primary
to
secondary
2.5KV/min Primary to safety ground 2.5KV/min Secondary to safety ground 2.5KV/min |
Insulation Resistance |
Primary
to
secondary
DC
2500V>500
megohm Primary to safety ground DC 2500V>500 megohm Secondary to safety ground DC 2500V>500 megohm |
Induction Withstand Voltage Test | 125Hz/800V/min |
Ground Resistance | 25A<100milliohm |
Cooling Method | Natural |
Noise Grade | <40dB |
Temperature Rise | < 115K |
Impedance Voltage | about 4% |
Total Losses | <5% |
Efficiency | >96% |
Operating Temperature | -25℃~40℃ (40℃ ,93% RH,56days) |
Storage Temperature | -25℃~100℃ (40℃ ,93% RH,56days) |
Certificate | CE,ISO |
The technical parameter for reference only, for exactly technical information, please feel free to contact us! |
Production process:
FAQ