Price | $12000 >=1 Set |
MOQ | 1 set |
Delivery Time | 35 work days |
Brand | Envsin |
Place of Origin | Guangdong |
Certification | ISO CE |
Model Number | EQF |
Payment Terms | L/C, D/A, T/T, D/P, Western Union |
Supply Ability | 20 sets per month |
Place of Origin | Guangdong | Warranty | 1 year |
Model Number | EQF | Supply Ability | 20 sets per month |
Certification | ISO CE | Brand Name | Envsin |
Payment Terms | L/C, D/A, T/T, D/P, Western Union | Gas concentration | 5ppm ~30ppm |
Wind speed | ≤ 1.7 m/s | Price | $12000 >=1 Set |
Delivery Time | 35 work days | Minimum Order Quantity | 1 set |
Gas | H2S,NO2,Cl2,SO2 | Power | AC380V 50Hz/60Hz 3/N/P |
NO2 Cl2 SO2 H2S Gas Corrosion Test Chamber 25ppm
Our service:
Main
Technical
Parameters
(Model:EQF)
|
||
Temperature test
|
Temperature
range: 20℃~+45℃
|
Remark:
1.
Test
in
accordance
with
IEC60068
under
empty
chamber
condition
(except
indicated).
2.
The
test
data
of
the chamber
are
under
the
conditions
that
the
water temperature
is
less
than
28,
the
water
pressure
is
between 0.3MPa
and 0.6MPa,
and
the water
pressure between
inlet
and
outlet
is
more
than
0.2MPa.
|
Temperature uniformity : ≤2.0℃ | ||
Temperature deviation: ≤±2.0℃ | ||
Temperature fluctuation: ≤ ±0.5℃ | ||
Temperature change rate: ≥ 1℃/min | ||
Temp
and
humidity test
|
Temperature range: 20℃~50℃ Humidity range: 50%RH~85%RH |
|
Humidity uniformity: ±5.0% RH | ||
Humidity deviation: +2.0% ~ 3%RH( >75%RH)±5%RH( ≤ 75%RH) | ||
Gas concentration | The deviation of gas concentration in the test chamber shall be less than 10% of the set concentration | |
The empty chamber operation is compliant with the IEC60068-2-60 | ||
Gas concentration measurement | calculate the volume ratio and then through the mass flowmeter measurement | |
On load capacity |
500W,The
temperature
can
be
stable,
the
chamber
body
is
not
deformed
|
|
Noise
|
Working noise ≤ 70dB ( measured at 1 meter away from the test chamber) | |
Gas flowmeter | H2S,CI2,SO2,NO2 | |
Standard
|
IEC60068-2-1:2007 Test A: Low temperature test method |
|
IEC60068-2-2:2007 Test
B:
High
temperature
test
method
|
||
IEC60068-2-78:2007 Test
Cab:
Constant
humidity
and