| Price | Negotiable |
| MOQ | 2000 |
| Delivery Time | 25-45 days |
| Brand | Zhaoli |
| Place of Origin | China |
| Certification | CE, ROHS, REACH |
| Model Number | DC2430PD01 |
| Packaging Details | cartons + pallets |
| Payment Terms | L/C, T/T, Western Union |
| Supply Ability | 10,000pcs per day |
| Printing | Customized printing | Place of Origin | China |
| Packaging Details | cartons + pallets | Model Number | DC2430PD01 |
| Supply Ability | 10,000pcs per day | Certification | CE, ROHS, REACH |
| Name | DC air pump | Bearing | sleeve bearing |
| Brand Name | Zhaoli | Payment Terms | L/C, T/T, Western Union |
| Packing | Customized packing | Delivery Time | 25-45 days |
| Rated voltage | 6.0VDC | Minimum Order Quantity | 2000 |
| Pump type | diaphragm air pump | Application | portable nebulizer/ blood pressure |
Diaphragm Air Pump DC Motor For Portable Atomizer Blood Pressure Supply With Electromagnetic Valve
Pump type: diaphragm pump
Specification:
| Test Environment | Ambient temperature of 20℃ and relative humidity of 65%RH. |
| Position | Horizontal |
| Load | Air pressure 300mmHg |
| Operating condition | -10℃~55℃, RH 40%~85% |
| Storage condition | -20℃~55℃, RH 5%~90% |
| Rated voltage | 6.0VDC |
| Operating voltage | 4.0 - 7.2 VDC |
| Rated current | The highest current is less than 380mA while pressurizing with DC6.0V from 0 to 300mmHg. |
| Inflation time | It takes Less than 10 seconds to pressurize with DC6.0V from 0 to 300 mmHg in a 500cc tank. |
| Leakage | Pressurizing from 0 till 300±30mmHg in a 500CC tank, then holding for 30 seconds. After that, measuring the dropped pressure value, which shall be less than 2.0mmHg/min. |
| Max pressure | The maximum pressure value shall be more than 400 mmHg while pressurizing with DC 6.0 V in a 500CC tank from 0 till dead-head pressure value. |
| Restart voltage | Fill pressure until 200mmHg pause,and then add DC 3.0V voltage pump should restart. |
| Dielectric strength | 600VAC/1S leakage current <=5mA |
| Temperature rise | <60K |
Drawing:

Application: This model is mainly applied in blood pressure or any other atomizer use.