China Maintaining Airflow and Pressure Control Laminar Flow Hood for sale
China Maintaining Airflow and Pressure Control Laminar Flow Hood for sale
  1. China Maintaining Airflow and Pressure Control Laminar Flow Hood for sale

Maintaining Airflow and Pressure Control Laminar Flow Hood

Price 1800$
MOQ 1
Delivery Time 14Days
Brand GCC
Place of Origin Guangdong,China
Certification ISO/GMP/CE
Model Number GCC-FG-1000
Packaging Details Wooden Box
Payment Terms T/T
Supply Ability 5000Sets/Month

Product Details

Product Specification

Usage Particle Filtration Place of Origin Guangdong,China
Packaging Details Wooden Box Material Stainless Steel
Model Number GCC-FG-1000 Supply Ability 5000Sets/Month
Feature Durability and Longevity Certification ISO/GMP/CE
Brand Name GCC Payment Terms T/T
Function Contamination Control Video outgoing-inspection Provided
Advantage Maintaining Airflow and Pressure Control Applicable Industries Manufacturing Plant
Price 1800$ Delivery Time 14Days
Product name Laminar Flow Hood Minimum Order Quantity 1

Product Description

Product Name: Laminar Flow Hood

Dimensions: 18009002600 (mm)

Cleanliness Level: Static Class 100 @ 0.3um (US Federal Standard)

Cabinet Material: SUS304 Stainless Steel

Average Face Velocity: 0.25~0.55m/s (adjustable in three levels), recommended 0.35m/s

HEPA Filters: 2 pcs, 78969369 (mm), efficiency 99.995%@≥0.3μm

Medium Efficiency Filters: 4 pcs, 75025021, efficiency 45%@≥0.5μm

Pre-filters: 4 pcs, 75025017, efficiency 40%@≥2.0μm

Noise Level: ≤65db

Power Supply: 220VAC, 50HZ, 550W (excluding power consumption of sockets)

Note: Pre-filters should be replaced every 1-3 months, medium efficiency filters every 3-6 months, and HEPA filters generally every year.

Replacement Guidelines:

  • HEPA filters must be replaced when the differential pressure reaches 400Pa.
  • Pre-filters must be replaced or cleaned when the differential pressure reaches 50Pa.
  • Medium efficiency filters must be replaced or cleaned when the differential pressure reaches 80Pa.

Key Features:

No Item Detail
1 Cleanliness Level Static Class 100 @ 0.3um (US Federal Standard)
2 Cabinet Material SUS304 Stainless Steel
3 Average Face Velocity 0.25~0.55m/s (adjustable in three levels), recommended 0.35m/s
4 HEPA Filters 2 pcs, 78969369 (mm), efficiency 99.995%@≥0.3μm
5 Medium Efficiency Filters 4 pcs, 75025021, efficiency 45%@≥0.5μm
6 Pre-filters 4 pcs, 75025017, efficiency 40%@≥2.0μm
7 Size(W*D*H)MM 1300*1200*600/1500*1200*600/2400*1200*600/2280*1460*500/1530*1060*500/1160*775*500
8 Note Pre-filters should be replaced every 1-3 months, medium efficiency filters every 3-6 months, and HEPA filters generally every year.

FAQ:

Q1:Are you a manufacturer or a trader? Where is your factory?
A1: We are a professional Purification Equipment manufacturer with a professional team of designers and engineers. Our factory is located in Guangzhou, China.

Q2:What is your delivery time?
A2: It will take 10-15 days after receiving your prepayment. The specific delivery time depends on your order quantity.